Mycoplasma Off™

36,90161,10

Clear
To order 1000ml Spray bottle outside of Germany the minimum required quantity is 5.
  • 100% Satisfaction Guaranteed
  • Questions? 1-908-524-4661 or Contact Us
  • Questions? +49 30 2000 4370 or Contact Us

Description

Mycoplasma Off™ is used for cleaning and disinfection of all kinds of laboratory surfaces and devices, including safety cabinets, incubators, work benches, sample storage boxes, and liquid nitrogen containers. Rapidly effective against mycoplasma contaminations, Mycoplasma Off™ has a broad antimicrobial activity against several other contaminants.

Benefits

  • Convenient and Rapid
  • Mycoplasma Off™ is a ready-to-use solution supplied in a refillable spray bottle. Mycoplasma Off™ is the ideal product to maintain a clean work area by ensuring fast and easy decontamination. Mycoplasma Off™ is immediately effective against mycoplasma (within few seconds after use).

  • Non-corrosive and Non-carcinogenic
  • Mycoplasma Off™ is an alcohol-based, non-corrosive and non-carcinogenic solution.

  • Broad Spectrum of Antimicrobial Activity
  • Bactericidal, fungicidal, and virucidal! Not only has a strong mycoplasmacidal activity but is also effective against Staphylococcus aureus, Enterococcus hirae, Escherichia coli, Proteus mirabilis, Pseudomonas aeruginosa, Candida albicans, BVD-, Polio, Adeno-, Polyoma SV40-, Papova-, MNV (Noro) and vaccinia viridae.

Recommended Use

Applicable in research and industry only. Not recommended for clinical applications. Use as directed. Mycoplasma Off™ should not be used on alcohol sensitive surfaces like acrylic and fiberglass. Always test a small area first to make sure the product does not cause damage or discoloration (e.g. on sensitive materials). For safe and convenient treatment of sensitive electric equipment, we also recommend our Mycoplasma Off™ Wipes.

Content

Ready-to-use formulation in a spray bottle or refill bottles.

Package Sizes
  • Cat. No. 15-1000 1 liter
  • Cat. No. 15-5000 5 liters canister
Required Consumables and Lab Devices

Paper towels

Shelf Life and Storage

Stable at room temperature for at least 12 months.

Additional information

We have a problem with our incubator. Something keeps growing inside our incubators. Can you give us suggestion how to clean and control this problem, please.

We like to suggest the following procedure:
1. Remove all racks and trays and clean them in a dish washer.
2. Switch off the incubator and let it cool down. Spray the inside of incubator wet with Mycoplasma Off or wipe it off with Mycoplasma Off Wipes or any other disinfectant containing alcohol >56 % (v/v) and glutaraldehyde. Let it air dry. Keep the door open for evaporation for at least 1 hour or preferably overnight. If your incubator is equipped with a heat inactivation procedure, start this procedure instead. Keep the racks and trays inside. Don´t forget to remove the CO2 sensor if necessary.
3. If your incubator is equipped with a copper inlay, do not polish the interior. Copper oxide is bacteriostatic. Polishing will remove this coating.
4. Reload the incubator. Use aqua dest. for the pan and add WaterShield.
5. For general operation, use plastic trays underneath your culture flasks to avoid spills during loading/unloading.
6. Use culture flasks with vent filters and keep the caps tight instead of standard flasks with open caps.